We have What you need for staining
AT-2000G GRAM
Stop hard Manual staining,Detail
Accurate and Prceison Stain
Can be used by simple operation and it can be available to obtain accurate and precision stain.Staining Results
Technical Specifications
AT-2000G GRAM Stainer | |
---|---|
Slide Tray Capacity |
1slide ~ 10slides (1 to 20 slides by adding Optional Tray) |
Tray Rotation Speed |
Stain Mode: Approx. 6RPM / Stain DRY Mode: Approx. 260RPM / DRY Mode: Approx. 200RPM |
Display | 4 Line by 20 character alpha-numeric LCD dot-matrix |
Key Button Controls |
6 Mode Key (Stain, Prime, Clean, Dry, Reset, Slide up/down) 6 Function selection |
Used Reagent / Consumption (Typical) |
Crystal Violet / Gram-lodine / Acetone Alcohol / Safranine 1.2ml ~ 2.2ml |
1 Cycle Stain time (Typical) |
10 slides: 10min / 20 slides: 13min. (option) |
Dimensions | 480(W) x 445(D) x 265(H) mm, Clearance Height(Lid open): Max. 565mm |
Weight | Approx. 17kgs |
Electrical Requirements |
100~240VAC~ (Auto selection)@50 to 60 Hz |
Power Consumption |
Max.50Watts |
Drain Connection |
Nipple on rear Panel |
Safety | Lid interlock: Lid must be closed, and is locked down during operation |
Standard Accessies |
5L Water bottle(1), Reagent Container(1), Reagent Container Cable(1), Power cord(1), Drain hose(1), Reagent connection Tube(1set), User’s manual(1), Quick intallation and user’s guide(1), Spare fuse(1), Carbon filter(5) |
Optional Accessaries |
Additional slide for 10 slides |